Full NIBP Determination Sequence for Adult
At each step the microprocessor stores cuff pressure, the
matched pulse amplitude, and the time between successive
pulses. The stepped deflation and matched pulse detection
continues until diastolic pressure is determined or total cuff
pressure falls below 8 mmHg. The parameter then deflates
the cuff (to zero detected pressure), analyzes the stored data,
and updates the screen.
The operating cycle is composed of four parts: inflation time,
deflation time, evaluation time, and wait time. Wait time,
which varies from mode to mode, is affected by the cycle
time (auto mode) or operator intervention (manual mode).
The figure shows the basic operating cycle for an NIBP
determination.