67
Using the Menu System
Response Mode (Nellcor Only)
Response mode allows the user to specify the averaging
technique to optimize measurements in the presence of
various patient movement. Choose Mode 1 (normal
Response) when patients are active as in exercise protocols.
Choose Mode 2 (Fast response; default setting) for the
general patient population.
SatSeconds™ (Nellcor Only)
The SatSeconds™ technique limit-controls the time that the %
SpO2 level may fall outside the alarm before an audible
alarm sounds. Choose either 0, 10, 25, 50, or 100 seconds.
If 0 is chosen this feature is disabled.
The SatSeconds™ “Safety Net” is for patients with saturation
levels having frequent excursions below the limit, but not
staying below the limit long enough for the SatSeconsd time
setting to be reached. When 3 or more limit violations occur
within 60 seconds, an alarm sounds even if the the
SatSeconds time setting has not been reached.
With Masimo SET technology installed, the Info button
displays the current settings for the Sensitivity and FastSAT
features. Sensitivity can be set to Normal (the factory
default) or Max. FastSAT can be set to On or Off (the factory
default).
Main
Selection of this button returns the user to the Main menu.
Config Button
Selection of this button displays the Config mode menu,
which allows the Power Save mode and time to be adjusted.