For monitors with Nellcor SpO
2
technology, the SpO
2
configuration menu includes Response Mode and
SatSeconds™. Response Mode can be set to “1” or “2”. The
factory default is “2”.
SatSeconds™ technique limit-controls the time that the %
SpO
2
level may fall outside the alarm before an audible
alarm sounds. Choose either 0, 10, 25, 50, or 100 seconds.
If 0 is chosen this feature is disabled
For monitors with Masimo SpO
2
technology, sensitivity can
be set to Normal (the factory default) or Max. FastSAT can
be set to On or Off.
Set BP. Selection of this button displays the message: Restore
BP mode on power-up? Selection of Yes restores the BP
mode to previous user-selected mode after power-off and
returns the Monitor to the More... menu. Selection of No
restores the BP mode to the default setting of Manual after
power-off and returns the Monitor to the More... menu.
Selection of Cancel returns the Monitor to the More... menu.
Alarms. Selection of this button displays the message: Enter
alarm configuration mode? Selection of No returns the
Monitor to the More... menu. Selection of Yes brings up the
Alarms menu. Selection of Reset changes all alarm limits
back to the factory defaults and returns the Monitor to the
More... menu. Selection of Save permanently saves the user-
selected alarm limits and returns the Monitor to the More...
menu. Selection of Cancel returns the Monitor to the More...
menu.
Main. Selection of this button returns the user to the Main
menu.