Using the Menu System
operating mode returns to the default setting of Manual. The
default setting of Manual can be overridden to return to the
previous user-selected setting (Auto or Manual) by selecting
Set BP under the Service menu. To cancel an Auto BP
determination, select the START/STOP key. To cancel an
Auto BP mode, select the Manual button in the Set BP (UK:
BP Mode) menu.
Manual
Selection of this option starts a Manual BP determination.
After power-off, the operating mode returns to the default
setting of Manual. The default setting of Manual can be
overridden to return to the previous user-selected setting
(Auto or Manual) by selecting Set BP under the Service menu.
Tgt Pressure
Selection of this option allows the user to set the BP target
inflation pressure. The initial target pressure can be set
between 100 and 250 mmHg in 5 mmHg increments. The
factory default is 150 mmHg for adults and 110 for neonates.
(This is indicated by “AUTO” at the end of adjustable range.)
When using a neonate blood pressure cuff, if the target
pressure is set to greater than 140 mmHg under the Set BP
or Clinical menu, the Monitor automatically defaults to a
target pressure of 110 mmHg. If the target pressure is set
between 100 and 140 mmHg, then that setting is the target
pressure that will be used. When the target pressure is
changed, the next determination will use the new target
inflation value if no systolic is available. Initial target pressure
is restored to the factory default setting after power-off. The
target pressure can be adjusted permanently in the Clinician
menu of the Service mode (refer to “Press” in the “Using the
Menu System” section).
Stat
Selection of this option allows the user to start Stat
determinations. When Stat is selected, blood pressure is
determined as many times as possible in 5 minutes.
Note: Alarm limits are disabled while in Stat mode.
Main
Selection of this button returns the user to the Main menu.