89
Appendix B
Procedural and Error Alarm Codes
Alarm
Code
LED
Display
LCD Description
Audible Tone
and Volume
Effect of
Alarm Silence
Switch
Effect of
Clear via
SelectKnob
Probable Cause
No
Code
No
change
LOW BATTERY,
Flashing
battery icon
3 beeps every
10 seconds,
adjustable
volume
2 minutes
silence
No effect
Replace or recharge battery.
From onset of alarm. 5 NIBP
measurements available.
Beep rate increases linearly
as battery discharges
No
Code
Blank
LOW BATTERY -
SYSTEM
DISABLED
Steady tone,
maximum
volume
No effect No effect
Replace or recharge battery.
NIBP measurement disabled
No
Code
No
change
PRINTER -
NO PAPER
High priority
alarm. Volume
adjustable
2 minutes
silence
Clear
Paper ran out or printer
door open
No
Code
Values
posted
NIBP RANGE
ERROR
High priority
alarm. Volume
adjustable
2 minutes
silence
Clear
NIBP algorithm returned
value outside specified
accuracy range
Other:
N, P,
E, I, S
Blank
Error code,
description
Steady tone,
maximum
volume
No effect No effect Internal system fault