When Auto mode is selected, a number at the right of the
Auto button indicates the time interval between each
reading. To change the time interval, choose the box around
the number and turn the rotor until the desired interval is
reached. The interval can be set between 1 and 120 minutes
(1, 2, 3, 4, 5, 10, 15, 20, 30, 45, 60, 90, and 120 minutes).
Press the rotor to confirm the setting. After power-off, the
operating mode returns to the default setting of Manual. The
default setting of Manual can be overridden to return to the
previous user-selected setting (Auto or Manual) by selecting
Set BP (UK: BP Mode) under the Service menu.
In the Auto mode, the pressure must be below 5 mmHg
(neonate) or 15 mmHg (adult) for at least 30 seconds before
another determination can be started. BP information will be
displayed on the LED until the next determination is started.
This applies to Auto mode only.
Note: To cancel an Auto BP determination, select the
Manual button in the Set BP menu (UK: BP Mode).
Stat Mode
Multiple BP readings can be taken at any time by selecting
the Stat button under the Set BP button (UK: BP Mode) in
the Main menu. Stat mode can also be accessed by pressing
the AUTO BP key (27) and then selecting the Stat button
when the Set BP menu (UK: BP Mode) appears.
If a Manual determination is not in progress, a 5-minute
series of determinations will start. If a Manual determination
is in progress, that determination will become the first in the
series. A normal, uninterrupted Stat sequence will give the
first set of systolic, diastolic, and mean arterial pressure
values and pulse rate within 15 to 20 seconds. Selecting the
Stat button during a series of Stat determinations will cancel
the determination in progress and the rest of the series. BP
information will be displayed on the LED until the
determination has been canceled or completed. This applies
to Stat mode only.
The series begins with cuff inflation to a pressure above the
previous systolic pressure or, if no previous systolic value is
stored, to approximately 150 mmHg for adult/pediatrics. The