Appendix A
Technical Specifications
BP
Cuff Pressure Range 0 to 290 mmHg (adult)
(Normal operating range) 0 to 140 mmHg (neonate)
Default Target: Cuff Inflation 150 ± 15 mmHg (adult)
110 ± 15 mmHg (neonate)
Target Cuff Inflation: 100 to 250 mmHg (adult)
Adjustment Range 100 to 140 mmHg (neonate)
(in 5 mmHg increments)
Blood Pressure Measurement
Range (mmHg) Systolic MAP Diastolic
Adult 30 - 245 15 - 215 10 -195
Neonate 40 - 140 30 - 115 20 -100
Blood Pressure Accuracy Meets or exceeds ANSI/AAMI
standard SP-10 (mean error 5
mmHg, standard deviation 8
mmHg)
Maximum Determination 120 s (adult)
Time 85 s (neonate)
Overpressure Cutoff 300 to 330 mmHg (adult)
150 to 165 mmHg (neonate)
Pulse Rate Range 30 to 200 beats/min (adult)
30 to 200 beats/min (neonate)
Pulse Rate Accuracy ± 3.5%
US Patents
4,360,029; 4,501,280; 4,546,775; 4,638,810; 5,052,397;
4,349,034; 4,543,962; 4,627,440; 4,754,761; 5,170,795;
5,518,000
European Patents
EP122123, EP205805, EP207807