Using the Monitor
initial target pressure selection for neonates is 110 mmHg.
Artifact rejection is relaxed in the Stat mode for adult/
pediatric patients to allow for accelerated determinations. If
a BP or Stat reading has been made previously, the first new
systolic value will flash on the LED display (7) within a few
seconds and will continue to flash until the end of the
determination. At that point a short tone will sound and the
updated systolic, diastolic, and mean arterial pressures and
pulse rate will appear on their LED displays (7, 9, 19, 22). The
Monitor will begin another determination once the pressure
is below 5 mmHg for 8 seconds (neonates) or 15 mmHg for
4 seconds (adults), unless the 5-minute period has ended or
the determination has been canceled.
Note: Alarm limits are disabled while in Stat mode.
TURBO
TEMP
Description
The Temperature parameter is included in Models 200V2
and 400V2. The PRO Monitor uses IVAC* TURBO
TEMP™
technology and can be used with both oral and rectal
temperature probes. The TURBO
TEMP™ parameter
consists of an electronic thermometer that uses a heat-
sensing device known as a thermistor to sense temperature.
The thermistor is part of the electrical circuit and is located at
the tip of the probe. To obtain temperatures, the probe tip
measures the rate of change in temperature when the
thermistor comes into contact with surrounding tissue. A
final temperature is calculated based on this rate of change.
During a temperature determination, the temperature display
(12) provides a progress meter and probe ready indicator. In
the far-left position, a single horizontal line indicates the
probe is ready to start a determination after removal from the
probe holster. In the far-right position of the temperature
display, a “chase sequence” around the outside space
indicates a predictive temperature determination is in
progress. During monitor mode, the temperature readings
flash constantly.