4-20 PRO 100-400 V2 Patient Monitor Service Manual Revision A
2009829-001
Calibration & Maintenance: SECTION 4. GENERAL MAINTENANCE
4.6.5. Adult Overpressure Check
1. Set up the unit and calibration equipment as shown in Figure 4-1.
2. Power on the PRO Monitor and select More...
3. Select the Service mode and input 2 2 1 3.
4. From the Service menu, select the NIBP button.
5. Select the Check button.
6. Increase applied pressure until overpressure occurs. Confirm that
pressure at the overpressure is between 300 mmHg and 330 mmHg
and the system pressure falls to less than 20 mmHg within 8 seconds.
7. Switch off the PRO Monitor and disconnect the calibration kit.
8. If the Monitor fails this test, re-calibrate the unit as described in
section 4.7.2.
4.6.6. Temperature System Check (PRO 200 & 400)
The PRO Monitor Series 200 & 400 temperature systems use
ALARIS Model 2885 and 2886 temperature probes. This system is
self-calibrating. The only maintenance required is to verify that the
temperature functions are working properly. These checks require an
IVAC probe simulator (P/N TE 1811), available from ALARIS Medical
Systems, Inc., San Diego, CA. GE Medical Systems Information
Technologies does not stock this tester.
To check the temperature system, connect the IVAC probe simulator
to the temperature probe connector on the front panel, and insert a
temperature probe into the active holster.
1. Power on the PRO Monitor.
2. Remove the temperature probe from the probe holster to initiate a
temperature reading.
Set the probe simulator to 98.6 and verify that the LCD temperature
display reads 98.6°F ±1.0°F. The numbers on the temperature LED
displays should be flashing at this point, indicating the monitor is in a
real-time monitor mode.
A range of temperatures can be checked, by using the other values
on the probe simulator (98.0, 80.2, 102.0, and 107.8).
Broken Probe Sensing
Set up the equipment and the probe simulator as in the previous
procedure. Rotate the temperature selector on the temperature
simulator to B.P. verify that the Temperature display reads 106.0°F