4-18 PRO 100-400 V2 Patient Monitor Service Manual Revision A
2009829-001
Calibration & Maintenance: SECTION 4. GENERAL MAINTENANCE
pressure value will be displayed in mmHg on the LCD. The system
will maintain the pressure for 36 seconds and will continue to display
the pressure in the pneumatic system on the LCD screen. Confirm
that the pressure has fallen no more than 10 mmHg while the system
is pressurized.
4.6.3. NIBP Calibration Check
This procedure verifies the linearity and calibration for both pressure
transducers (PT1 and PT2) over the range 0-250 mmHg.
To verify calibration it is necessary to have the following test
equipment:
- 2-tube NIBP hose, 12ft, p/n 107365
- DINAMAP Adult BP cuff p/n 2774 (or equivalent)
- Calibration Kit p/n 320246 with a manual inflation bulb
- NIST calibrated single-tube manometer
Figure 4-1
1. Set up the unit and calibration equipment as shown in Figure 4-1.
Vista de pagina 69