Revision A PRO 100-400 V2 Patient Monitor Service Manual 4-13
2009829-001
Calibration & Maintenance: SECTION 4. GENERAL MAINTENANCE
NIBP Button
The NIBP button accesses the NIBP Service menu. Use this menu to
calibrate and check the NIBP transducers and the overpressure
circuitry (Section 4.5.3).
Info Button
This button displays the date of the most recent NIBP calibration.
Select OK to return the display to the Service menu.
System Button
This button accesses the System menu. The options on this menu set
the display language, communications protocols, test the EEPROM,
and print the error log (Section 4.5.4).
Silence Button
Selecting the Silence button displays a request to confirm the choice.
This is a working option available only through the Clinician Mode
(1234). Alarms are disabled when in Service Mode.
Yes mutes all patient alarms until the PRO Monitor is powered off and
on again or the alarm Silence button is selected again to enable the
alarms. A confirmation menu appears (pictured above) on the display.
Selecting either Yes or No exits the menu. If silence is confirmed, the
alarm Silence button on the front panel illuminates momentarily.
4.5.3. NIBP Menu
The options on the NIBP menu, pictured above, are used to calibrate
and test BP functions of the PRO Monitor.