Revision A PRO 100-400 V2 Patient Monitor Service Manual 4-11
2009829-001
Calibration & Maintenance: SECTION 4. GENERAL MAINTENANCE
Select Reset to return all the alarm limits to the default settings. The
illustration above shows the default settings.
The user can adjust the alarm limits within the ranges specified in the
table below. The high limit must be at least one step higher than the
low limit.
Note: Predictive temperature has no alarm limit alarm.
Select Save to save the displayed settings and return to the More…
menu. Select Cancel to abandon any changes and return to the
More… menu.
4.5. SERVICE MODE
The Service Mode provides the same features as the Clinician Mode
and adds these functions:
Check the BP calibration
Re-calibrate BP
Adjust the pneumatic system overpressure point
Configure communications with a host computer
Change the language of operation
4.5.1 Accessing the Service Mode
To enter Service Mode from the Main menu, use the rotor and LCD.
1. Select the More… button.