Revision A PRO 100-400 V2 Patient Monitor Service Manual 4-15
2009829-001
Calibration & Maintenance: SECTION 4. GENERAL MAINTENANCE
case, the Monitor prompts the user to select a language on every
power up until a selection occurs.
Comms Button
This button accesses the Communications menu. The settings on this
menu configure communications with a host computer when PRO
Monitors are connected to a network.
The Remote Op check box toggles remote operation. When on, a
check mark (√) appears in the box. When on, the PRO Monitor
responds to external commands and can initiate a BP determination
when prompted by a host computer. Remote operation requires
DINAMAP
®
Host Communications Protocol, which is described in the
DINAMAP
®
Host Communications Reference Manual.
The Standard check box sets the host comms protocol to standard
format, and the baud rate at 9600 bps.
The Address box sets the Monitor address within a range of 32 to
126. Address 32 is not unique. All monitors, regardless of unit
address, respond to commands to unit address 32.
Selecting the 1846 button sets the comms protocol to 1846 format,
and the baud rate to 600 bps.
The baud Rate can be set to 300, 600, 1200, 2400, 4800, 9600, or
19200.
Select OK to return to the Main Service menu.
eprom Button (EEPROM Testing)
This button initiates a test of the read and write functions on the
EEPROM. The test is repeated 20 times. The results are displayed on
the LCD and recorded in the error log.
Err Log Button
This button generates a printout of the last 20 system errors that have
occurred on the unit. The system and secondary processor software
versions are also listed on the error log.